Magnetic Microscope
Site:
Hamburg
RESOLUTION
(nm)
-
300.0
-
SAMPLE SIZE
(mm)
35.0
Magneto-Optical Domain Imaging and Magnetometry
Applications:
Measurement of magnetic hysteresis of thin films, micro- and nanoelements.
Magnetic domain imaging of thin films, micro- and nanoelemente
Optics
Reflection geometry
Objectives with magnification: 20x, 50x, and 100x
Optical resolution: 300 nm
Magneto-optical Kerr effect: Polar, longitudinal and transversal MOKE contrast
Sample Environment
Typical sample sizes < 35 mm
Piezo-based drift correction to improve magnetic domain imaging
Magnetic Field Options
Quadrupole magnet for in-plane magnetic fields up to 700 mT depending on sample size
Perpendicular coil for out-of-plane magnetic fields up to 100 mT
DC and AC magnetic field modes
Post Processing
Software for magnetic hysteresis curve evaluation